Casio GRAPH 25+ Pro, GRAPH 35+, GRAPH 75, GRAPH 85, GRAPH 85SD, GRAPH 95 Instrukcja Użytkownika Strona 366

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 413
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 365
20051101
4-2
Using a Custom Probe
6. Use the function keys described below to configure the custom probe setup.
To change the setting of an item, first use the f and c cursor keys to move the
highlighting to the item. Next, use the function keys to select the setting you want.
(1) Slope
Press 1(EDIT) to input the slope for the linear interpolation formula.
(2) Intercept
Press 1(EDIT) to input the intercept for the linear interpolation formula.
(3) Unit Name
Press 1(EDIT) to input up to eight characters for the unit name.
(4) Warm-up
Press 1(EDIT) to input the warm-up time.
7. Press wand then input a memory number (1 to 99).
This saves the custom probe setup and returns to the Custom Probe List, which should
now contain the new custom probe setup you configured.
uu
uu
u
To recall the specifications of a Vernier sensor and configure custom
probe settings
1. Perform the first two steps of the procedure under “To configure a custom probe setup” on
page 4-1.
2. Press 5(VRNR).
This displays a Vernier sensor list.
3. Use the f and c keys to move the highlighting to the Vernier sensor whose setting
you want to use as the basis of the custom probe settings, and then press w.
The name and specifications of the Vernier sensor you select will appear on the custom
probe setup screen.
To complete this procedure, perform steps 4 through 7 under “To configure a custom
probe setup” (page 4-1).
Przeglądanie stron 365
1 2 ... 361 362 363 364 365 366 367 368 369 370 371 ... 412 413

Komentarze do niniejszej Instrukcji

Brak uwag